Assetize

All learning series · Manage capability · Part 1

Calibration

Keep calibration results, tolerance decisions and follow-up work connected.

Start here

Confirm this product is ready, then use the lesson. The readiness checks are the gate for configuration, for a first journey and for production.

  1. Are you ready to use Calibration? Licensing and AppPoints, the records that must exist, and three checklists.
  2. Basics
  3. Configuration
  4. Practical example
  5. Practice and verify
  6. Advanced concepts

Are you ready to use Calibration?

Calibration records what an instrument read, against a tolerance, on a data sheet that you can trace. It is included with Manage. Have the instrument, the points and the tolerance before you run the first sheet.

Before configuration

Manage entitlement that includes Calibration, a site, and a decision about the standard and the unit each point uses.

Before the first successful journey

One instrument asset, one data sheet with at least one test point and a tolerance, and a technician who can enter an as-found reading.

Before production use

Agree who owns the standards, how often sheets are revised, and how a failed point becomes work. History is the point of the product, so the first production week needs a place those results will be kept.

Licensing, entitlement and AppPoints

Calibration is entitled and installed as part of Maximo Manage. The MAS 9.2 licensing guidance includes Calibration at the Base and Limited tiers and does not list a separate Calibration install quantity. A Limited user counts calibration applications inside the three modules they may choose. Reference standards and any laboratory accreditation are your own obligations. They are not an AppPoint line.

Keep four facts separate for Calibration. Entitlement is the proof of entitlement and the License Information for IBM Maximo Application Suite. The MAS 9.2 licensing guidance supplements that agreement. Available AppPoints are the pool shown on the AppPoints Usage dashboard: Suite administration, then Suite, then Administration, then AppPoints Usage. Assigned access is the security group together with the user type on the person: Limited, Base or Premium, and Authorized or Concurrent. Consumption is what that assignment draws. Authorized users and administrators reserve AppPoints continuously. Concurrent users draw AppPoints while they are signed in. Install AppPoints stay reserved for the life of that environment. A non-production install of a component draws no install AppPoints when the same component is already installed in production. Administrators and User Management users open the dashboard. Other people need the LICENSEUSAGE security group. Compliance enforcement and idle timeout sit on the dashboard Configuration tab and are suite-administrator settings.

The same guidance publishes these user conversion ratios: Limited is 5 AppPoints for one concurrent user and 2 for one authorized user; Base is 10 concurrent and 3 authorized; Premium is 15 concurrent and 5 authorized; an administrative Base user is 10 AppPoints and an administrative Premium user is 15, both as authorized users. A Base administrator who already reserves 10 AppPoints can be given Premium application access with no further AppPoints. Those ratios were read from the MAS 9.2 licensing guidance on 18 October 2026. Your agreement sets how many AppPoints you purchased. An application that opens shows that the component is deployed. Read the proof of entitlement and the dashboard as well.

Installed and enabled applications

Calibration applications are part of Manage. Confirm data sheets and the calibration work path open. Installed means the applications are on the workspace. Configured means a data sheet has points, units and tolerances. Usable means a technician can enter a reading and the sheet shows whether it met the tolerance.

Foundation configuration

You need an asset that is the instrument, a classification or item description that identifies it, and the unit of measure for each test point. A location tells you where the instrument sits. A job plan is how you repeat the sheet. It can come after the first manual sheet. Storerooms and certificates matter when you issue a replacement instrument. They are not required to record one as-found and one as-left reading.

Product configuration

Create the data sheet, add the test points, the nominal value, the unit and the tolerance. Associate it with the instrument. Grant the calibration applications and work-order access to the technician group. Review the default that treats a point as passed when the tolerance is still empty, and put a tolerance on every point you will use in production.

Infrastructure and integration

Calibration runs inside Manage. No external service is required for the sheet itself. A laboratory system or a document store for certificates is a separate integration and needs its own access and licence.

Data and operational readiness

The first journey needs one instrument, one sheet, one point with a unit and a tolerance, and two readings: one inside the tolerance and one outside it, so you can see both results. A missing tolerance is a configuration gap. A missing instrument is data. A hidden application is security.

Setup order for a calibration

  1. 1. Manage entitlement that includes Calibration No separate install quantity.
  2. 2. Instrument asset, site and unit The thing you will calibrate, and the unit the point uses.
  3. 3. Data sheet, point and tolerance The sheet is the procedure. The tolerance is the decision.
  4. 4. As-found and as-left readings One result inside the tolerance and one outside it.
Readiness for Calibration
RequirementWhy neededMandatory/Conditional/OptionalWhere to configure or verifyMinimum working exampleVerification procedureConsequence if missingRemediationSource
Manage entitlement including CalibrationCalibration is part of Manage at Limited and Base, with no install line of its own.MandatorySuite administration → Suite → Administration → AppPoints Usage. Manage → Security → Security Groups, and the suite user type.A technician as Base, or Limited with calibration inside the three modules.Open the data-sheet application and read the user’s tier.The calibration applications are missing from the menu.Grant the applications. Do not look for a Calibration install charge on the dashboard.Licensing in Maximo Application Suite 9.2
An instrument asset in a siteThe sheet is about a specific instrument.MandatoryAssets, in the site where the work will be recorded.Asset INST-1, site training, description pressure gauge.Query the asset and confirm status allows work.The sheet has no asset to attach to.Create the asset in the site before you build the sheet.Modules and applications in Maximo Application Suite 9.2
A data sheet with a point, a unit and a toleranceWithout a tolerance the reading has nothing to pass or fail against.MandatoryData sheets, points, nominal, unit, tolerance.Point 1, nominal 100 kPa, tolerance ±2 kPa.Save the sheet and read the point back, including the unit and both tolerance bounds.Technicians enter numbers that the sheet cannot judge.Add the unit and the tolerance before the first live reading.Modules and applications in Maximo Application Suite 9.2
A technician security groupEntering a reading and approving a sheet are different from viewing an asset.MandatoryManage → Security → Security Groups, and the suite user typeGroup CAL-TECH with the calibration applications and work-order save on the training site.Sign in as the technician and open the sheet. Sign in as a viewer and confirm the sheet is read only.The technician sees the sheet and cannot save a reading.Grant the signature option that saves calibration results.User access and entitlements in MAS 9.1 and later
One in-tolerance and one out-of-tolerance reading on a training copyThe journey is only real when both outcomes are visible.MandatoryThe calibration work record for INST-1.As-found 101 kPa (passes) on one point, and 110 kPa (fails) on a second training point or a second pass.Save both and read the pass and fail indicators back.Every point looks the same and the tolerance is untested.Enter the out-of-tolerance reading on a training sheet before you use the sheet on a production instrument.Modules and applications in Maximo Application Suite 9.2
A standard or certificate storeTraceability to a reference standard is how a regulated site defends the result.ConditionalThe standard record or the attached certificate, when your procedure requires one.Standard STD-1, due date after the test date, linked from the sheet.Open the sheet and navigate to the standard. Confirm the due date is still valid on the test date.The reading is stored and the standard it depended on is unknown.Link the standard before production use. A training journey can record the reading first.Licensing in Maximo Application Suite 9.2

A. Ready to configure

  • The user’s tier includes Calibration and the dashboard agrees.
  • The training site and instrument asset exist.
  • The technician group can open the calibration applications.
  • You know the unit and the tolerance the first point will use.

B. Ready for the first end-to-end journey

  • The data sheet is saved with at least one point, a unit and a tolerance.
  • A technician enters an as-found reading inside the tolerance and the sheet shows a pass.
  • A reading outside the tolerance shows a fail on that point.
  • Both results read back on the instrument’s calibration history.

C. Ready for production operation

  • Every production point has a unit and a tolerance. Empty tolerances are reviewed before go-live.
  • Standards have owners and due dates, where your procedure uses them.
  • A failed point has an agreed next step, such as a work order or a quarantine status.
  • Someone other than the technician can read the history for an instrument.

Sources for this readiness check

Licence statements come from the IBM topic named on the row. Installation of a component is a separate check from entitlement.

Before you start

Calibration is a Manage capability. Use the installed calibration guide and locally approved measurement procedure for application execution.

  • Calibration capability with training instrument/data-sheet access.
  • A fictional measurement range, agreed tolerances and test results.

IBM sources + Assetize exercise

01 · Basics — Understand the building blocks

Calibration capability is included in Manage. Calibration work needs a defined measurement procedure and an explicit interpretation of the observed result.

As-found and as-left results answer different questions: what condition the instrument arrived in and what condition it was left in.

Topics in this series

  • Instruments and data sheets
  • Test points and tolerances
  • As-found results
  • Adjustment and as-left results
  • History and follow-up

02 · Configuration — Work through the setup

Use a training environment and match the actions to the IBM guide for your installed build.

  1. Select a training instrument and inspect its calibration requirements.
  2. Define or select an approved training data sheet with test points and tolerances.
  3. Create the associated calibration work through the installed process.
  4. Record as-found values, then any adjustment and as-left values according to that procedure.
  5. Review the result, follow-up requirements and retained history.

03 · Practical example

Fictional teaching example

In a teaching example a 100-unit test point allows 99 to 101. An as-found value of 103 is outside the example tolerance; a later as-left value of 100 does not erase the original result.

04 · Practice and verify

Record one passing and one out-of-tolerance test and explain the required follow-up.

Expected result

Original and final results remain distinguishable, with a traceable tolerance decision.

Common mistakes

  • Replacing an as-found failure with an as-left pass.
  • Using undocumented tolerances.

05 · Advanced concepts

Explore reference standards, overdue calibration and assessment of prior measurements after an out-of-tolerance result.